Exfo MAX-710-P Manuel d'utilisateur Page 363

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 381
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 362
Description of Event Types
MAX-700 351
Merged Event
The total loss ( dB) produced by the events is measured by plotting
two straight lines.
The first line is plotted by fitting, through least-square
approximation, trace points in the linear area preceding the first
event.
The second line is plotted by fitting, through least-square
approximation, trace points in the linear area following the second
event. If there were more than two merged events, this line would
be plotted in the linear area following the last merged event. This
line is then projected toward the first merged event.
The total loss ( dB) equals the power difference between the
point where the first event begins (point A) and the point on the
projected straight line located just below the first event (point B).
No loss value can be specified for the subevents.
Vue de la page 362
1 2 ... 358 359 360 361 362 363 364 365 366 367 368 ... 380 381

Commentaires sur ces manuels

Pas de commentaire