Operating the Polarization Mode Dispersion Analyzer
38 FTB-5500B
Setting Up Hardware for Measurements
5. If applicable, select the C bandwidth.
6. If the FLS-110 Light Source supply indicator flashes, press auto off to
deactivate the Auto Off mode. If this mode remains enabled, the light
source could stop emitting during the test.
7. Turn the source on.
8. Press high to select the highest output power. The HIGH LED lights up.
Note: The low output power can be used but the accuracy of the PMD will be
affected.
Note: For more information on the FLS-110P, please refer to the FLS-110 Light
Source user guide included with your unit.