Exfo FTB-5600 Distributed PMD Analyzer Manuel d'utilisateur Page 104

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 146
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 103
Maintenance
98 FTB-5600
Recalibrating the Unit
Recalibrating the Unit
Manufacturing and service center calibrations are based on the
ISO/IEC 17025 Standard, which states that calibration documents must not
contain a recommended calibration interval, unless this has been
previously agreed upon with the customer.
Validity of specifications depends on operating conditions. For example,
the calibration validity period can be longer or shorter depending on the
intensity of use, environmental conditions and unit maintenance. You
should determine the adequate calibration interval for your unit according
to your accuracy requirements.
Under normal use, EXFO recommends calibrating your unit every year.
Vue de la page 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 145 146

Commentaires sur ces manuels

Pas de commentaire