Optimizing Measurements
126 FTB-5600
Use an Appropriate Residual PMD for the Acquisition
The figure below shows an example of a link having a low PMD measured
with bad PMD scale and high PMD resolution.
A better setting for this acquisition would have been a PMD scale of 6 ps
and a residual PMD value of 133 fs.
Resulting
PMD
curve
Vue de la page 131