Configuring the FTB-7000D OTDR on the IQS-12002B
94 IQS-12002B
Creating the GP-121310: Calibration Instrument for Singlemode, Front-Connector Position
4. From the View Standard Parameters window, click Edit to enter the
length given on the acceptance test plan for all requested wavelengths.
5. Click Close.
IMPORTANT
Do not change the Length_Tolerance parameter.
IMPORTANT
Now that you have created a calibration instrument for
UPC connectors, you must manually create a calibration instrument
for APC connectors. For more information, see Creating the
GP-121310: Calibration Instrument for Singlemode,
Front-Connector Position (APC Connectors) on page 98.