Exfo MaxTester Series Manuel d'utilisateur Page 132

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 175
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 131
Maintenance
122 MaxTester Series
Cleaning Detector Ports
Cleaning Detector Ports
Regular cleaning of detectors will help maintain measurement accuracy.
To clean detector ports:
1. Remove the protective cap and adapter (FOA) from the detector.
2. If the detector is dusty, blow dry with compressed air.
3. Being careful not to touch the soft end of the swab, moisten a cleaning
tip with only one drop of isopropyl alcohol.
4. While applying light pressure (to avoid breaking the detector window),
gently rotate the cleaning tip on the detector window.
5. Repeat step 4 with a dry cleaning tip or blow dry with compressed air.
6. Discard the cleaning tips after one use.
IMPORTANT
Always cover detectors with protective caps when unit is not in use.
IMPORTANT
Alcohol may leave traces if used abundantly. Do not use bottles that
distribute too much alcohol at a time.
Vue de la page 131
1 2 ... 127 128 129 130 131 132 133 134 135 136 137 ... 174 175

Commentaires sur ces manuels

Pas de commentaire