Setting up Your Fiber Inspection Probe and ConnectorMax2
Fiber Inspection Probe 27
Managing and Selecting Test Configurations
7. If you want to modify the information about the inspection zones,
proceed as follows:
7a. Tap the button corresponding to the desired inspection zones.
7b. Modify the parameters as needed to indicate whether you want to
be notified of scratches, defects or both for each zone in the fiber,
then set thresholds for each item you select.
You can set up to 3 criteria per zone, and per anomaly type
(scratches or defects). The thresholds are divided into three
categories:
Any: this enables the next criterion, which requires a specific
value.
1 to 10: the next criterion is automatically filled out to show the
infinity symbol ( ) and 0 as a threshold.
0: the criterion definition is complete.