Exfo FTB-3930 MultiTest Manuel d'utilisateur Page 39

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 101
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 38
Measuring Optical Return Loss
Multifunction Loss Tester 33
Performing and Saving ORL Measurements
5. Connect one end of a patchcord to the FASTEST SM port of your unit
leaving the other end unconnected.
6. Perform an ORL reference as follows:
Note: During the reference, the patchcord end should remain in the air;
reflections occurring at a fiber-to-air interface correspond to a constant of
14.7 dB.
Press ORL Reference.
Vue de la page 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 100 101

Commentaires sur ces manuels

Pas de commentaire