SCPI-Befehlsreferenz
OSA 499
Product-Specific Commands—Description
INBAND: InBand noise type is selected.
INBANDNARROWFILTER: InBand narrow filter
noise type is selected.
POLYNOMIAL5: 5th order polynomial fit noise
type is selected.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "ITU_011", 201.4670 THZ
CALC:WDM:CHAN:SEL "ITU_011"
CALC:WDM:CHAN:NOIS:AUTO ON
<Do measurement>
CALC:WDM:DATA:CHAN:SEL "ITU_011"
CALC:WDM:DATA:CHAN:NOIS:TYPE? Returns
INBAND or INBANDNARROWFILTER
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:TYPE
:CALCulate[1..n][:WDM]:DATA:CHANnel:NOISe?
:CALCulate[1..n][:WDM]:DATA:CHANnel:NOISe:
AUTO?
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:
NOISe:TYPE?
Vue de la page 506