Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manuel d'utilisateur Page 595

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 653
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 594
SCPI-Befehlsreferenz
OSA 587
Product-Specific Commands—Description
:STATus?
Description This query returns a value indicating the global
status of the instrument.
This command has no associated *RST
condition.
Syntax :STATus?
Parameter(s) None
Response Syntax <Status>
Response(s) Status:
The response data syntax for <Status> is defined
as a <CHARACTER RESPONSE DATA> element.
The <Status> response represents the
instrument state, where:
UNINITIALIZED means the instrument has not
been initialized yet.
INITINPROGRESS means the instruments
initialization is in progress.
READY means the instrument is ready.
BUSY means the instrument is busy.
DISCONNECTED means the instrument is
disconnected.
DEFECTIVE means the instrument is defective.
Example(s) STAT?
See Also :CALibration:ZERO:AUTO?
:INITiate[:IMMediate]
:INITiate:CONTinuous?
:STATus:OPERation:BIT<8|9>:CONDition?
Vue de la page 594
1 2 ... 590 591 592 593 594 595 596 597 598 599 600 ... 652 653

Commentaires sur ces manuels

Pas de commentaire