Exfo FTB-5700 Single Ended Dispersion Analyzer for FTB- Manuel d'utilisateur Page 72

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 94
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 71
Troubleshooting
66 FTB-5700
Solving Common Problems
Unable to find a reflective
fiber end. Please verify that
the fiber under test is
properly connected and
terminated by a UPC
connector.
Too much loss in the fiber
under test
³ Clean the fiber end.
³ Add a reflective
termination at the end
of the fiber.
Optical power is too low. Too much loss in the fiber
under test.
³ Clean the fiber end.
³ Add a reflective
termination at the end
of the fiber.
Unable to find a valid
wavelength range.
³ Too much loss in the
fiber under test.
³ The wavelength range is
too large.
³ Clean the fiber end.
³ Add a reflective
termination at the end
of the fiber.
³ Reduce the wavelength
range in the Acquisition
tab of the Setup
window, or select
Default as the
wavelength range value.
Message Possible cause Solution