SCPI Commands Reference
Optical Modulation Analyzer 201
Product-Specific Commands—Description
:SENSe:DDEMod:WINDow:AMPLitude?
Description This query returns the measurement value for
amplitude thresholds.
At *RST value, it is set to 20%-80%.
Syntax :SENSe:DDEMod:WINDow:AMPLitude?
Parameter(s) None
Response Syntax <amplitude>
Response(s) amplitude:
The response data syntax for <amplitude> is
defined as a <CHARACTER RESPONSE DATA>
element.
Returns the measurement value for amplitude
thresholds.
20-80, 20%-80% is selected as the measurement
value for amplitude thresholds.
10-90, 10%-90% is selected as the measurement
value for amplitude thresholds.
Example(s) LINS1:SENS:DDEM:WIND:AMPL?
See Also SENS:DDEM:WIND:LOW?