Exfo FastReporter 2 Manuel d'utilisateur Page 351

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 396
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 350
Description of Event Types
Post Processing Software 341
Continuous Fiber
Continuous Fiber
This event indicates that the selected acquisition range was shorter than
the fiber length.
The fiber end was not detected because the analysis process ended
before reaching the end of the fiber.
The acquisition distance range should therefore be increased to a
value greater than the fiber length.
There is no loss or reflectance specified for continuous fiber events.
Reflected power
(dB)
Continuous
fiber
Distance
(km)
Vue de la page 350
1 2 ... 346 347 348 349 350 351 352 353 354 355 356 ... 395 396

Commentaires sur ces manuels

Pas de commentaire