Exfo FastReporter 2 Manuel d'utilisateur Page 353

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 396
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 352
Description of Event Types
Post Processing Software 343
Non-Reflective Event
Non-Reflective Event
This event is characterized by a sudden decrease in the Rayleigh
backscatter signal level. It appears as a discontinuity in the downward
slope of the trace signal.
This event is often caused by splices, macrobends, or microbends in
the fiber.
A loss value is specified for non-reflective events. There is no
reflectance specified for this type of event.
If you set thresholds, the application indicates a non-reflective fault in
the events table, whenever a value exceeds the loss threshold.
Reflected power
(dB)
Distance
(km)
Non-reflective
event
Linear downward slope
due to Rayleigh
backscatter
Vue de la page 352
1 2 ... 348 349 350 351 352 353 354 355 356 357 358 ... 395 396

Commentaires sur ces manuels

Pas de commentaire