Remote Control
94 FLS-2600B
Specific Commands
OUTPut:STATe?
Description This query returns the current source state.
Syntax OUTP:STAT?
Response A boolean value representing the current source state:
0: the source is deactivated.
1: the source is activated.
See also OUTP:STAT, OUTP:MODE, OUTP:MODE? and OUTP:ASE?
SOURce:MODE
Description This command is used to toggle between the normal and
high-resolution modes.
Syntax SOUR:MODE<wsp><mode>
Parameters The <mode> parameter is a boolean value representing the
mode state:
0: Normal mode
1: High-Resolution mode
Example SOUR:MODE 1
Note This command cannot be processed if your tunable laser
source is performing a sweep.
See also SOUR:MODE?