Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manuel d'utilisateur Page 415

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 688
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 414
Справка по командам SCPI
OSA 409
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Distance> parameter.
The <Distance> parameter corresponds to a valid
distance in hertz from peak to center of noise
region.
The
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency? MIN and
CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency? MAX queries can be
used to determine valid distance values.
Example(s) CALC:WDM:CHAN:AUTO:NOIS:TYPE POLY5
CALC:WDM:CHAN:AUTO:NOIS:DIST:FREQ
100.0 GHZ
CALC:WDM:CHAN:AUTO:NOIS:DIST:FREQ?
Returns 1.000000E+011
Notes Custom noise measurement distance is applied
only if selected noise type is POLYnomial5.
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance[:WAVelength]
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
DISTance:FREQuency?
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
WIDTh:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:AUTO:NOISe:
TYPE
:CALCulate[1..n][:WDM]:CHANnel:NOISe:DISTa
nce:FREQuency
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
NOISe:DISTance:FREQuency
Vue de la page 414
1 2 ... 410 411 412 413 414 415 416 417 418 419 420 ... 687 688

Commentaires sur ces manuels

Pas de commentaire