Справка по командам SCPI
428 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
SIGnalPower:TYPE
Description This command selects WDM analysis default
channel's signal power measurement type.
At *RST, the signal power type is set to IPOWer
(integrated power).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:SIGnal
Power:TYPE<wsp>IPOWer|PPOWer|TPOWer
Parameter(s) Type:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
IPOWer|PPOWer|TPOWer.
The parameter corresponds to the newly selected
signal power type.
IPOWer: selects integrated signal power type.
PPOWer: selects peak signal power type.
TPOWer: selects channel total power type.