OSA 635
Product-Specific Commands—Description
:TRIGger[1..n][:SEQuence]:SOURce?
Description This query returns the selected the source for the
trigger system event detector.
At *RST, the source is set to IMMediate.
Syntax :TRIGger[1..n][:SEQuence]:SOURce?
Parameter(s) None
Response Syntax <Source>
Response(s) Source:
The response data syntax for <Source> is defined
as a <CHARACTER RESPONSE DATA>
element.
The <Source> response corresponds to the
selected trigger event source.
IMMediate: No waiting for an event occurs.
TIMer: The source signal comes from a periodic
timer.
Example(s) TRIG:SOUR IMM
TRIG:SOUR? Returns IMMEDIATE
See Also :INITiate[:IMMediate]
:INITiate:CONTinuous
:TRIGger[1..n][:SEQuence]:SOURce