Exfo FTB -5240S/5240BP Optical Spectrum Analyzer for FT Manuel d'utilisateur Page 459

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 537
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 458
SCPI Command Reference
Optical Spectrum Analyzer 451
Product-Specific Commands—Description
:CALCulate[1..n]:CHANnel:COUNt?
Description This query returns the number of channels on
the list.
*RST clears the channel list.
Syntax :CALCulate[1..n]:CHANnel:COUNt?
Parameter(s) None
Response Syntax <ChannelCount>
Response(s) ChannelCount:
The response data syntax for <ChannelCount>
is defined as a <NR1 NUMERIC RESPONSE
DATA> element.
Number of channels on the list.
Example(s) CALCulate:CHANnel:COUNt?
CALCulate:CHANnel:ADD C34,1550.235
NM,50.1G,Hz,INTEGRATEDSIGNAL,FIXEDIEC
See Also CALCulate[1..n]:CHANnel:ADD