SCPI Command Reference
496 FTB-5240S/5240BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:COUNt
Description Sets the number of scans that will be averaged to
produce final trace.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 8.
Syntax :SENSe[1..n]:AVERage:COUNt<wsp><AverageC
ount>
Parameter(s) AverageCount:
The program data syntax for <AverageCount> is
defined as a <numeric_value> element.
Selects the number of scans that will be
averaged to produce final trace.
Range: [2 ... 9999]
Example(s) SENS:AVER:TYPE AVERAGING
SENS:AVER:COUN 12
INITiate:IMMediate
See Also SENSe[1..n]:AVERage:COUNt?
SENSe[1..n]:AVERage:TYPE