SCPI Command Reference
Optical Spectrum Analyzer 467
Product-Specific Commands—Description
:CALCulate[1..n]:POWer:OFFSet
[:OUTPut]
Description This command sets the Power Offset value.
For this command to be accepted, the OSA
module must be in Ready state with no
acquisition in progress.
At *RST, this value is 0.00 dB.
Syntax :CALCulate[1..n]:POWer:OFFSet[:OUTPut]<wsp
><Offset>
Parameter(s) Offset:
The program data syntax for <Offset> is defined
as a <DECIMAL NUMERIC PROGRAM DATA>
element.
Selects the Offset value.
Range: [-99.99 … 99.99] dB
Example(s) CALC:POWer:OFFSet -0.56
See Also CALCulate[1..n]:POWer:OFFSet:OUTPut?