Vue de la page 118
SCPI Command Reference
Variable Attenuator 113
Product-Specific Commands—Description
:INPut[1..n]:WAVelength?
Description This query returns a value indicating either the
current or the minimum/maximum wavelength.
At *RST, the wavelength that will be selected
depends on the instrument you have.
Syntax :INPut[1..n]:WAVelength?[<wsp>MAXimum|MI
Nimum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <Wavelength>