SCPI Command Reference
Variable Attenuator 139
Product-Specific Commands—Description
:OUTPut[1..n]:RPOWer?
Description This query returns a value indicating either the
current or the minimum/maximum relative
power value.
At *RST, the value that will be set depends on the
instrument you have.
Syntax :OUTPut[1..n]:RPOWer?[<wsp>MAXimum|MINi
mum|DEFault]
Parameter(s) Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
lowest supported value.
MAXimum is used to retrieve the instrument's
highest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax <RelativePower>
Response(s) RelativePower:
The response data syntax for <RelativePower>
is defined as a <NR3 NUMERIC RESPONSE
DATA> element.
The <RelativePower> parameter corresponds to
a valid relative power value.