SCPI Command Reference
142 IQS-3150
Product-Specific Commands—Description
:OUTPut[1..n][:STATe]?
Description This query returns the state of the instrument's
shutter.
At *RST, the state of the shutter is set to off
(closed).
Syntax :OUTPut[1..n][:STATe]?
Parameter(s) None
Response Syntax <ShutterState>
Response(s) ShutterState:
The response data syntax for <ShutterState> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <ShutterState> represents the current
shutters state.
0: the shutter is closed (no light is transmitted).
1: the shutter is open (allows light transmission).
Example(s) OUTP:STAT ON
OUTP:STAT? Returns: 1 (the shutter is open)
See Also :OUTPut[1..n][:STATe]
:OUTPut[1..n]:LOCK[:STATe]?