Referencia de instrucciones SCPI
418 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel
[:DEFine]
Description This command allocates and initializes a new
WDM analysis channel setup.
*RST has no effect on this command.
Syntax :CALCulate[1..n][:WDM]:CHANnel[:DEFine]<ws
p><Name>,<Define[<wsp>M|HZ]>|MAXimu
m|MINimum
Parameter(s)
Name:
The program data syntax for <Name> is defined
as a <STRING PROGRAM DATA> element.
The <Name> parameter corresponds to the
name of the new channel setup to create. The
channel name cannot be empty.
Each channel name must be unique: it is not
possible to define two channels with the same
name.
Define:
The program data syntax for <Define> is defined
as a <numeric_value> element followed by an
optional <SUFFIX PROGRAM DATA> element.
The allowed <SUFFIX PROGRAM DATA>
elements are: M|HZ. The <Define> special
forms MINimum and MAXimum are accepted on
input.