Referencia de instrucciones SCPI
558 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:SENSe[1..n]:AVERage:COUNt
Description This command sets the number of acquired
traces to combine for averaging to a specific
value.
At *RST, averaging count is set to 8.
Syntax :SENSe[1..n]:AVERage:COUNt<wsp><Count>|
MAXimum|MINimum|DEFault
Parameter(s) Count:
The program data syntax for <Count> is defined
as a <numeric_value> element. The <Count>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Count> parameter.
The <Count> parameter corresponds to a valid
averaging count value.
The SENSe[1..n]:AVERage:COUNt? MIN and
SENSe[1..n]:AVERage:COUNt? MAX queries can
be used to determine valid count range.