Exfo FTB-5230S Optical Spectrum Analyzer for FTB 200 v2 Manuel d'utilisateur Page 530

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 659
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 529
Referencia de instrucciones SCPI
524 FTB-5240S/S-P/BP
Product-Specific Commands—Description
DEFault allows the instrument to select a value
for the <Resolution> parameter.
The <Resolution> parameter corresponds to the
custom resolution bandwidth in meter.
The
CALCulate[1..n][:WDM]:OSNR:BANDwidth[RESo
lution]? MIN and
CALCulate[1..n][:WDM]:OSNR:BANDwidth[RESo
lution]? MAX queries can be used to determine
valid resolution bandwidth range.
Example(s) CALC:WDM:OSNR:BAND:RES:AUTO OFF
CALC:WDM:OSNR:BAND:RES 0.065 NM
CALC:WDM:OSNR:BAND:RES? Returns
6.500000E-011
See Also :CALCulate[1..n][:WDM]:BWIDth[1|2]|BANDwi
dth[1|2]:RelativeLEVel
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]?
:CALCulate[1..n][:WDM]:OSNR:BWIDth|BANDw
idth[:RESolution]:AUTO
:CALCulate[1..n][:WDM]:THReshold
:CALCulate[1..n][:WDM]:OSNR:
BANDwidth|BWIDth[:RESolution]
Vue de la page 529
1 2 ... 525 526 527 528 529 530 531 532 533 534 535 ... 658 659

Commentaires sur ces manuels

Pas de commentaire