Referencia de instrucciones SCPI
OSA 421
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel
[:DEFine]?
Description This query requests the instrument to return the
definition of the specified WDM channel analysis
setup.
*RST has no effect on this command.
Syntax :CALCulate[1..n][:WDM]:CHANnel[:DEFine]?<w
sp><Name>
Parameter(s) Name:
The program data syntax for <Name> is defined
as a <STRING PROGRAM DATA> element.
The <Name> parameter corresponds to the
name of the channel setup definition to request.
Response Syntax <Define>