Exfo FTB-7000 OTDR Series for FTB-500 Manuel d'utilisateur Page 449

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 482
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 448
SCPI Command Reference
OTDR 435
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:AM:INTernal:
FREQuency
Description This command selects the internal modulation
frequency of the visual fault locator (VFL). The
internal modulation corresponds to 50 % of the
duty cycle at the selected frequency.
*RST sets the modulation frequency to 0 Hz
(CW).
Syntax :SOURce[1..n]:VFLocator:AM:INTernal:FREQuen
cy<wsp><Frequency>|MAXimum|MINimum|
DEFault
Parameter(s) Frequency:
The program data syntax for <Frequency> is
defined as a <numeric_value> element. The
<Frequency> special forms MINimum,
MAXimum and DEFault are accepted on input.
Vue de la page 448
1 2 ... 444 445 446 447 448 449 450 451 452 453 454 ... 481 482

Commentaires sur ces manuels

Pas de commentaire