SCPI Command Reference
442 OTDR
Product-Specific Commands—Description
:SOURce[1..n]:VFLocator:POWer:
STATe?
Description This query returns a value indicating if the visual
fault locator (VFL) is on or off.
*RST sets the VFL to OFF.
Syntax :SOURce[1..n]:VFLocator:POWer:STATe?
Parameter(s) None
Response Syntax <State>
Response(s) State:
The response data syntax for <State> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
Power state of the VFL (on or off).
Example(s) SOUR:VFL:POW:STAT ON
SOUR:VFL:POW:STAT? Returns 1
See Also SOURce[1..n]:VFLocator:POWer:STATe
SOURce[1..n]:VFLocator:POWer:STATe:TIME
SOURce[1..n]:VFLocator:AM:STATe?
SOURce[1..n]:VFLocator:AM:INTernal:FREQuenc
y