Test Results
NetBlazer 329
Service Configuration - Ramp
Service Configuration - Ramp
From the Test menu, tap Results, Service Configuration, and the Ramp
tab.
Service Name and Selection
Service Name indicates the name of the selected service. Select the
service to be displayed by either using the left/right arrow or by tapping
over the service numbers area then tapping on a specific service number.
An orange background indicates the selected service while a green
background indicates the services that are enabled.
Committed/Excess Steps
Committed Steps indicate the pre CIR and CIR steps specified in the
ramp configuration.
Excess Steps indicate the CIR+EIR and Traffic Policing steps
specified in the ramp configuration.
Direction, available with Dual Test Set, indicates results from local to
remote (L -> R) and remote to local (R -> L).
TX Rate indicates the transmission rate for each step.
SLA Verified indicates the committed SLA parameters that are used to
declare the pass/fail verdict. See Summary (EtherSAM) on page 347 for
more information on Frame Loss, Max Jitter, Round-Trip Latency,
and Max RX Rate.
Informational parameters are for information purpose only, they are
not included in the test pass/fail verdict. See Summary (EtherSAM) on
page 347 for more information on Frame Loss, Max Jitter, and
Round-Trip Latency.
Average RX Rate indicates the measured average utilization
throughput for each step.