Test Results
NetBlazer 345
Summary (Cable Test)
Wire Map Test Result gives the wire map test result for each pair.
When the link is up: The wire map result for each pair is given as seen
by the FTB-700G/800 Series to get a link up. This means that the wire
map result may not correspond to the type of cable tested depending
on the cable(s) used and/or the configuration of the cable mode (MDI,
MDIX, or auto-detection) on both the FTB-700G/800 Series and the far
end equipment. For example, two crossed pair cables end to end used
between the FTB-700G/800 Series and a far end equipment may give a
straight pair (MDI) wire map result.
Note: For 1Gbit/s, both MDI and MDIX can be reported simultaneously since
crossed pairs detection is performed independently for pairs A-B and C-D.
MDI Straight pair.
MDIX Crossed pair.
MDI (-) For 1Gbit/s, straight pair with swapped wires within pair.
MDIX (-) For 1Gbit/s, crossed pair with pair A swapped with pair
B and/or pair C swapped with pair D.
Noise Excessive noise on a pair most likely caused by a link
partner running in 10/100 Mbit/s forced mode. In this
case, no propagation delay or length is reported and
there is no comparison with any threshold.