SCPI Command Reference
Tunable Laser Source 105
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:STOP
Description This command allows you to specify a sweep
stop wavelength setting.
At *RST, the stop wavelength is 1570 nm.
Syntax :SOURce[1..n]:WAVelength:SWEep:STOP<wsp
><StopWavelength[<wsp>M|HZ]>|MAXimum
|MINimum|DEFault
Parameter(s) StopWavelength:
The program data syntax for <StopWavelength>
is defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: M|HZ. The
<StopWavelength> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <StopWavelength> parameter.
The <StopWavelength> parameter is the desired
stop spectrum value for the sweep. If you do not
specify a unit, the device will use the current
spectral units (meter or Hz). To select the current
spectral units, use the UNIT[1..n]:SPECtrum
command.