SCPI Command Reference
120 IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
DWELl
Description This command allows you to set the time interval
between each step of the Step-by-Step sweep
mode.
At *RST, the time interval between each step is 2
seconds.
Syntax :SOURce[1..n]:WAVelength:SWEep:DWELl<wsp
><DwellTime[<wsp>S]>|MAXimum|MINimu
m|DEFault
Parameter(s) DwellTime:
The program data syntax for <DwellTime> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> element is S. The
<DwellTime> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <DwellTime> parameter.
This parameter is the duration of the pause
between the steps in seconds.