SCPI Command Reference
84 IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength[:CW]
Description This command selects the source's output
wavelength (in meters) or frequency (in Hz).
At *RST, the wavelength is device-dependent.
Syntax :SOURce[1..n]:WAVelength[:CW]<wsp><Wave
length[<wsp>M|HZ]>|MAXimum|MINimum|D
EFault
Parameter(s) Wavelength:
The program data syntax for <Wavelength> is
defined as a <numeric_value> element
followed by an optional <SUFFIX PROGRAM
DATA> element. The allowed <SUFFIX
PROGRAM DATA> elements are: M|HZ. The
<Wavelength> special forms MINimum,
MAXimum and DEFault are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Wavelength> parameter.
The parameter represents the spectrum value in
meters or in hertz.
DEF = 1550 nm.