Exfo IQS-2600B Tunable Laser Source for IQS Platforms Manuel d'utilisateur Page 115

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 142
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 114
SCPI Command Reference
Tunable Laser Source 109
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:SWEep:
SPEed
Description This command allows you to select the sweep
speed for continuous mode.
At *RST, the sweep speed is 15 nm/s.
Syntax :SOURce[1..n]:WAVelength:SWEep:SPEed<wsp
><Speed>|MAXimum|MINimum|DEFault
Parameter(s) Speed:
The program data syntax for <Speed> is defined
as a <numeric_value> element. The <Speed>
special forms MINimum, MAXimum and DEFault
are accepted on input.
MINimum allows to set the instrument to the
smallest supported value.
MAXimum allows to set the instrument to the
greatest supported value.
DEFault allows the instrument to select a value
for the <Speed> parameter.
This parameter sets the optical spectrum
displacement sweep speed in meters/s.
Example(s) SOUR:WAV:SWE:SPE 35.455E-9 (in meters/s)
Notes This function cannot be performed while the
module is sweeping.
See Also SOURce[1..n]:WAVelength:SWEep:SPEed?