SCPI Command Reference
88 IQS-2600B
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:MODE
Description This command is used to toggle between normal
(NORMal) and High Resolution (HRESolution)
modes.
At *RST, resolution mode is NORMal.
Syntax :SOURce[1..n]:WAVelength:MODE<wsp>NORM
al|HRESolution
Parameter(s) SourceMode:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
NORMal|HRESolution.
This parameter is the sources new resolution
mode.
NORMal sets the source to normal operation
mode.
HRESolution sets the source to high resolution
operation mode.
Example(s) SOUR:WAV:MODE HRES
See Also SOURce[1..n]:WAVelength:MODE?