Exfo MaxTester MAX-610 Manuel d'utilisateur Page 273

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 281
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 272
MaxTester Copper Test 265
A Technical Specifications
IMPORTANT
The following technical specifications can change without notice.
The information presented in this section is provided as a reference
only. To obtain this product’s most recent technical specifications,
visit the EXFO Web site at www.exfo.com
COPPER SPECIFICATIONS
a, b, c
Transmitter characteristics
Frequency range (200 Hz to 20 kHz) Frequency resolution 1-Hz steps
Frequency uncertainty (accuracy) ±(50 ppm + 1 Hz)
Level range (dBm) –20 to 10 at 600 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 600
Frequency range (20 kHz to 2.2 MHz) Frequency resolution 1-kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range (dBm) –20 to 10 at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150
Frequency range (2.2 MHz to 30 MHz) Frequency resolution 1-kHz steps
Frequency uncertainty (accuracy) ±(50 ppm + 100 Hz)
Level range (dBm) –20 to 0 at 100 1
Level resolution 0.1 dB
Level uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150
Receiver characteristics Reception frequency range 200 Hz to 20 kHz
20 kHz to 30 MHz
Frequency uncertainty range (accuracy) ±(50 ppm + 1 digit)
VF reception level range (dBm) –90 to 15 at 600 1
VF level uncertainty (accuracy) 200 Hz to 20 kHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
WB reception level range (dBm) –90 to 15 at 100 1 and 135 1
WB level uncertainty (accuracy)
20 kHz to 2.2 MHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
2.2 MHz to 30 MHz
–90 dBm to –50 dBm, uncertainty (accuracy) ±2 dB
–50 dBm to 15 dBm, uncertainty (accuracy) ±1 dB
Impedance (1) 100, 120, 135, 150, 600
POTS dialer DTMF 0 – 9, #, *
Phonebook 25 entries
Digital multimeter (DMM) Test type Snapshot and continuous
Impedance selection (for voltage measurement) 100 k1, 1 M1
Measurement Range Resolution Uncertainty (accuracy)
DC voltage 0 to 400 V 0.1 V for 0 to 99.9 V
1 V for 100 V to 400 V
±(|1 %| + 0.5 VDC)
AC voltage 0 to 280 Vrms 0.1 V for 0 to 99.9 V
1 V for 100 V to 280 V
±(1 % + 0.5 VAC)
Isolation resistance
(stress/leakage)
0 to 1 G1, auto-ranging
1 k1 to 99 M1
100 M1 to 999 M1
Three digits
±(2 % + 1 digit)
±(5 % + 1 digit)
Resistance 0 to 100 M1
0 to 999 1
1 k1 to 100 M1
Three digits
±(1 % + 5 1)
±(2 % + 1 digit)
Capacitance 0.1 nF to 2 μF Four digits ±(2 % + 50 pF)
DC current 0 to 110 mA 0.1 mA ±(|2 %| + 1 mA)
AC current 0 to 110 mA 0.1 mA ±(|2 %|+ 1 mA)
d
Station Ground 0 to 1 M1
0 to 999 1
1 k1 to 1 M1
Up to three digits
±(1 % + 3 1)
±(2 % + 1 digit)
NOTES
a. Subject to change without notice.
b. Typical, at 23 °C ± 3 °C, on batteries, with no type B USB connection.
c. Specifications based on 24 AWG (PE 0.5 mm) cabling.
d. From 10 mA to 110 mA.
Vue de la page 272

Commentaires sur ces manuels

Pas de commentaire