SCPI Command Reference
358 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:CHANnel:AUTO:
CENTer:ITUGrid?
Description This query indicates if snap center on ITU grid for
WDM analysis default channel has been enabled
or not.
At *RST, snap center on ITU grid is set to off
(disabled).
Syntax :CALCulate[1..n][:WDM]:CHANnel:AUTO:CENTer
:ITUGrid?
Parameter(s) None
Response Syntax <Auto>
Response(s) Auto:
The response data syntax for <Auto> is defined
as a <NR1 NUMERIC RESPONSE DATA>
element.
The <Auto> response corresponds to the state
of the snap center on ITU grid.
0: snap center on ITU grid is disabled.
1: snap center on ITU grid is enabled.
Example(s) CALC:CHAN:AUTO:CENT:ITUG OFF
CALC:CHAN:AUTO:CENT:ITUG? Returns: 0 (snap
ITU grid disabled)
See Also :CALCulate[1..n][:WDM]:CHANnel:AUTO
:CALCulate[1..n][:WDM]:CHANnel:AUTO:CENTer
:ITUGrid