SCPI Command Reference
534 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:MEMory:TABLe:SELect?
Description This query returns the name of the currently
selected table.
At *RST, there is no selection: a single null string
is returned.
Syntax :MEMory:TABLe:SELect?
Parameter(s) None
Response Syntax <TableName>
Response(s) TableName:
The response data syntax for <TableName> is
defined as a <STRING RESPONSE DATA>
element.
The <TableName> response corresponds to the
name of the currently selected table.
Example(s) MEM:TABL:SEL "WDM:CHANNEL"
MEM:TABL:DEF "NAME,CMAS:WAV"
MEM:TABL:SEL? returns "WDM:CHANNEL"
Notes The only valid table name is "WDM:Channel".
See Also :MEMory:TABLe:DEFine
:MEMory:TABLe:DEFine?
:MEMory:TABLe:SELect