SCPI Command Reference
490 FTB-5240S/S-P/BP
Product-Specific Commands—Description
:CALCulate[1..n][:WDM]:DATA:CHANnel:
SELect
Description This command sets the name of the selected
WDM channel result.
At *RST, this value is not available.
Syntax :CALCulate[1..n][:WDM]:DATA:CHANnel:SELect
<wsp><Select>
Parameter(s) Select:
The program data syntax for <Select> is defined
as a <STRING PROGRAM DATA> element.
The <Select> parameter corresponds to the
name of the channel result to select. The channel
name cannot be empty.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:DEF "C_007", 1380.000 NM
<Do measurement>
CALC:WDM:DATA:CHAN:SEL "C_001"
See Also :CALCulate[1..n][:WDM]:CHANnel:SELect
:CALCulate[1..n][:WDM]:DATA:CHANnel:CATalo
g?
:CALCulate[1..n][:WDM]:DATA:CHANnel:NSELec
t
:CALCulate[1..n][:WDM]:DATA:CHANnel:SELect?