Exfo FTB -5240S/5240BP Optical Spectrum Analyzer Manuel d'utilisateur Page 605

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 638
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 604
SCPI Command Reference
OSA 599
Examples on Using the SCPI Commands
Performing a Single Acquisition
Command Sequence Comments
<Add commands to configure analysis
parameters>
SENS:CORR:OFFS:MAGN 5.0 DB Set power offset.
SENS:WAV:OFFS 0.065 NM Set wavelength offset.
SENS:WAV:STAR 1525.000 NM Set sweep wavelength range: 1525.000 nm to
1570.000 nm.
SENS:WAV:STOP 1570.000 NM ""
SENS:AVER:STAT OFF Disable trace averaging.
TRIG:SEQ:SOUR IMM Set sweep trigger event source to immediate.
STAT? Test instrument state is idle. Poll STAT? until the
returned state is READY.
INIT:IMM Start sweep acquisition.
STAT:OPER:BIT8:COND? Wait for the acquisition to be completed. Poll
bit 8 until the returned value is 0.
<Add commands to retrieve analysis
results>
Vue de la page 604
1 2 ... 600 601 602 603 604 605 606 607 608 609 610 ... 637 638

Commentaires sur ces manuels

Pas de commentaire