SCPI Command Reference
OSA 523
Product-Specific Commands—Description
With <Continuous> set to OFF, the trigger
system remain in idle state until <Continuous>
is set to ON or INITiate:IMMediate command is
received. With <Continuous> set to ON, the
trigger system leave idle state and continue
cycling until <Continuous> set to OFF or ABORt
command is received.
When <Continuous> is set to OFF, the current
trigger cycle is completed before returning to idle
state: current acquisition continue until finished.
Example(s) CALC:WDM:CHAN:DEL:ALL
CALC:WDM:CHAN:AUTO ON
CALC:WDM:CHAN:AUTO:NOIS:AUTO OFF
CALC:WDM:CHAN:AUTO:NOIS:TYPE IEC
CALC:WDM:STATe ON
TRACe:FEED:CONTrol "TRC1", ALW
SENS:AVER:STAT OFF
:INITiate:CONTinuous