Exfo IQS-12004B DWDM Passive Component Test System for Manuel d'utilisateur Page 180

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 437
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 179
Automating or Remotely Controlling the System
168 IQS-12004B
A typical test sequence would be similar to the flow chart below.
Each time a new test sequence is started, the reference step is highly
recommended.
Each step is further explained in the following pages.
Calibrate system
IL/PDL
ORL
Reference system
IL/PDL
ORL
Test DUT
Optimize power
Initialize hardware
Null power meter
Select input port
Measure/analyze
Retrieve data
C
y
c
l
i
n
g
u
s
i
n
g
t
h
e
s
w
i
t
c
h
Te s t DUT
Retrieve results
Standard IQS-12004B
DWDM Passive
Component Test System
(one DUT at a time)
MultiPath Testing
option (possibility of
several DUTs at a time)
Vue de la page 179
1 2 ... 175 176 177 178 179 180 181 182 183 184 185 ... 436 437

Commentaires sur ces manuels

Pas de commentaire