Automating or Remotely Controlling the System
DWDM Passive Component Test System 189
GPIB Specific Commands—Quick Reference
ACQ? (0 | 1) The response to this query indicates
if a wavelength response calibration
has been performed.
NULL -- This command performs a nulling
on the power meter and wavelength
reference module channels.
ORLR -- This command performs an ORL
calibration with a reference test
jumper.
CAL ORLR? (0 | 1) The response to this query indicates
if an ORL calibration with a
reference test jumper has been
performed.
STOP -- This command stops the current
calibration acquisition.
ORLM -- This command performs an ORL
calibration acquisition with mandrel
and reference test jumper.
ORLM? (0 | 1) The response to this query indicates
if an ORL calibration acquisition
with mandrel and reference test
jumper has been performed.
REF ORLM -- This command performs an ORL
reference acquisition using a
mandrel.
ORLM? (0 | 1) The response to this query indicates
if an ORL reference using a mandrel
has been performed.
STOP -- This command stops the current
reference acquisition.
Command Parameter/Response Description