COM Objects Reference
276 IQS-12004B
Measures Interface—Methods
Measures
AcqPowerSynchronized
Description This command performs a synchronized power acquisition on
all IQS-1600 power meters.
Syntax object.AcqPowerSynchronized (NumberOfPorts,
NumberOfAcquisitionPoints, FireEvent)
Parameters Byval Long NumberOfPorts
Byval Double NumberOfAcquisitionPoints
Byval Boolean FireEvent
Response None
Possible error ErrInitBefore, ErrHardNotCompatible, ErrCalWrBefore,
ErrRefIlPdlBefore, ErrInProgress, ErrAcqType,
ErrNumberOfAcquisitionPoints,
ErrNotAvailableInSimulatorMode, ErrAcqStopped,
ErrScanRange, ErrSourceNotEnabled. For more information,
see IQS-12004B DWDM Passive Component Test System
(Standard) COM Error Messages on page 208.
Note The maximum number of points is 32 000.