Exfo IQS-12004B DWDM Passive Component Test System for Manuel d'utilisateur Page 423

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 437
  • Table des matières
  • DEPANNAGE
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 422
Definitions and Calculation Methods
DWDM Passive Component Test System 411
Flatness/Ripple (dB)
The flatness (or ripple) is the IL variation over the BW1 interval of each
channel.
Note: The ripple value for a measured BW1 analysis would most often be the
power (dB) value at which BW1 is measured.
Ripple
Ripple
1.0 dB cutoff wavelengths
Vue de la page 422