Specifications
1
Loss measurement uncertainty
2, 3, 5
± 0.06 dB
Loss measurement repeatability
2, 3, 6
± 0.04 dB
PDL uncertainty
2, 4, 5
± (0.05 dB + 5 % of DUT PDL)
PDL repeatability
2, 4, 6
± (0.025 dB + 3 % of DUT PDL)
ORL uncertainty
2, 5, 7
± 1 dB (0 dB to 45 dB)
± 2 dB (45 dB to 50 dB)
ORL repeatability
2, 3, 6
± 0.3 dB (0 dB to 45 dB)
ORL range
2
0 dB to 55 dB
Operating temperature Any stable temperature at ± 1 °C
in the range 20 °C to 30 °C
All other specifications Please refer to IQS-12004B data sheet
Standard Accessories
Instruction manual, installation software CD
and interconnecting patchcords
Optional Accessories
Switch module with launch fibers
Rackmount kit
Video monitor
Notes:
1. All specifications are typical, after 60-minute warmup and testing a low
loss components using the recommended EXFO accessories and EXFO
low PDL switch.
2. Using a semi-rigid patchcord between the IQS-9401 and the switch.
3. Valid for spectrally uniform component. Does not include uncertainties
due to connector or connector adapter.
4. Valid for spectrally uniform component terminated with a non-angled
connector. Does not include uncertainties due to connector or
connector adapter.
5. For first 100 cycles, within 100 hours after calibration and reference.
6. For 100 consecutive cycles within 100 hours.
7. Calibrated using a UPC connector with 14.7 dB of return loss.
Commentaires sur ces manuels