Operating the Optical Power Meter
Optical Power Meter 23
Programming an Acquisition
Programming an Acquisition
You can program the PM-1100 to automatically acquire absolute or
referenced measurements. The following program parameters may be set:
³ delay: the beginning of the program may be delayed by up to
999 hours, 59 minutes, and 59 seconds;
³ duration: the program can last up to 999 hours, 59 minutes, and
59 seconds;
³ number of samples: up to 1024 samples can be taken (depending on
the selected duration).