Remote Control Commands
Optical Power Meter 85
Product-Specific Commands—Description
MMEMory:ACQuisition:DELete
Description This command clears the 1024 memory locations reserved for
the acquisition data.
Syntax MMEM:ACQ:DEL
MMEMory:ACQuisition:DELAy
Description This command changes the delay parameter for the acquisition
setup.
Syntax MMEM:ACQ:DELA<space><delay>
Parameters The <delay> parameter is the delay before the acquisition
starts in the “HHH,MM,SS” format (the maximum delay is
999 hours, 59 minutes, 59 seconds).
Example MMEM:ACQ:DELA 000,01,50
MMEMory:ACQuisition:DURAtion
Description This command changes the duration parameter for the
acquisition setup.
Syntax MMEM:ACQ:DURA<space><duration>
Parameters The <duration> parameter is the duration of the acquisition
setup in the “HHH,MM,SS” format (the maximum duration is
999 hours, 59 minutes, 59 seconds).
Example MMEM:ACQ:DURA 000,01,50